Water Analyzer PH ORP MLSS DO......

  • Multi-Parameter Analyzer
Multi-Parameter Analyzer

Multi-Parameter Analyzer





This Multi-Parameter could monitor the PH/ORP Conductive and Temperature at the same time, it was connect with 2 electrode, one is work for PH or ORP(Oxidative redox potential), the other is work for Conductive and Temperature. It could work for Power Station, Petrochemical, Metallurgy, Mining industry, Paper industry, Fermentation Engineering, Medicine, Food & Beverage , Environmentally friendly water treatment, Aquaculture,Modern farming. CE Certification

   Character:
   * Flexibility, reliability and low maintenance
   * Temperature compensation
   * 4 alarm relay
   * 2 groups output
   * RS485//RS232 Communication Interface Optional
   * Menu operation mode
   * 128*64 Large LCD display

Specification:
1. Measuring Range:
* pH: 0 ~14.00pH
* ORP: -1999mV ~ +1999mV
* Conductivity: 0 ~ 100mS/cm
* Temperature: -5C ~ 110.0C

2. The Basic Error:
* PH: ±0.01PH;
* ORP: ±2mV
* Conductivity: ±0.5%FS
* Temperature: ±0.3C

3. Current Output: 0~10mA or 4~20mA
4. 4 Alarm Relay: 3A 240VAC (Max)
5. Power Consumption: 220V~240VAC <3W; 110V<2W; 24VDC<1W;
6. Optional: RS485//RS232 Communication Interface

PH/ORP  Electrode:
* PHD-01: 1-14PH 0~80C glass electrode for sewage, industy water, river water
* PHD-02: 1-14PH 0~80C glass electrode for pure water, Boiler water
* PHD-03: 1-14PH 0~80C Antimony electrode for Hydrofluoric water
* PHD-04: 1-14PH 0~130C PTFE glass electrode for High Temperature water
* PHD-05: 1-14PH 0~130C Easyferm Plus 120 electrode for Fermentation & Medicine

Conductivity Electrode:
* CED-01: 0 ~ 20μS/cm K=0.01 0~60C
* CED-02: 0 ~ 200μS/cm K=0.1 0~60C
* CED-03: 0 ~ 20μS/cm K=0.01 0~150C
* CED-04: 0 ~ 200μS/cm K=0.1 0~150C
* CED-05: 0 ~ 1000μS/cm K=1.0 0~60C
* CED-06: 0 ~ 20mS/cm K=10.0 0~60C
* CED-07: 0 ~ 100mS/cm K=30 0~150C

CONTACT US

Contact: Alvin

Tel: 400-1199-703

Email: sales@ayite.net

Add: No.116-122, KWOK SHUI ROAD, Kwai Chung Industry Center, HongKong